Accurate yet
Affordable Constant Current
The 2500EP and
the Higher impedance version (2500EN) constitute
Valhallas most popular current calibrator. The ranges of
the 2500EP and 2500EN are compatible with virtually all
analog meters and most of the digital multimeters with
AC-DC current capability. For benchtop DMMs and the
handheld meters, virtually all with 10 amp measurement,
the 10 amp range of the 2500EP is ideal. Ammeters,
current transformers, shunts, current sensors,
coils,wattmeters and a variety of parametric tests
requiring constant, yet variable, AC or DC current are
but a few of calibration and lab test uses of the 2500EP.
The units fast response characteristics allow it to
precisely track the input reference voltage in terms of
amplitude, phase, frequency, and waveshape. Each range
also features zero to 200% dynamic response capability
(up to maximum output).
10 Volt
Compliance Drive
Compliance
voltage to a current calibrator is what drive current is
to a voltage source. It is the ability to tolerate high
series resistance (up to 10 V drop) and still deliver
stable, constant current. The 2500 series current
calibrators all feature good drive compliance capability
to deliver constant current long after competitive units
drop out. A built in compliance indicator informs the
user of compliance status at all times.
Remote Ranging
- Systems Interfaceable
For automated
use, the 2500EP has remote ranging. The remote range is
selected by applying plus five volts to the proper rear
panel connector pin. For GPIB, this signal is available
from the Valhalla 2701C DC Voltage Standard. The 2500EP
has front panel L.E.D. remote range indicators and a
rotary knob for manually setting the desired
range.
|
- Dynamic
Range: 0 to 200% of range, DC or RMS AC up to maximum
rated output (12 Amps maximum)
- DC Accuracy:
(180 days, 25°C ±5°C) ±0.01% of
output ±0.02% of Full Scale
- AC Accuracy:
(180 days, 25°C ±5°C) ±0.05% of
output ±0.1% of Full Scale
- Input
Impedance: 100 Kilohms, except the 2500EN is 10
Megohms
- Compliance
Voltage: 10 Volts DC or pk AC
- Input/Output
Ratio: 2.00000 volt input provides full- scale output,
20 volts for 2500EN (200% of range)
- Stability:
±0.0005%/15 min. - 0.005% for 24
Hours
- Output
Protection: Fuse protected at 15 amps Max.Input: 3
VDC/pk AC, 30 VDC/pk AC for 2500EN
- Response
Time: 1 millisecond to ±0.01% of final value
following input amplitude or frequency
change
- Maximum
Isolation Voltage: ±200 VDC or peak
AC
- Input Common
Mode Rejection: 60db @ DC linearly decreasing to 40 db
@ maximum frequency
- Temperature
Coefficient: ±0.001% of output ±0.002% of
range/°C below 20°C or above 30°C; double
the above for AC
- Power:
115/230 VAC ±10% 50-60Hz
- Size:
381mm/15" L x 432mm/17" W x 133mm/5.23" H
- Weight: 14
Kg/30 lbs. Net 18 Kg/40 lbs. shipping
weight
|