The 2701C-1PPM 24
Hr. Stability 8 PPM 30 Day Accuracy
Valhalla
Scientific's 2701C Programmable Precision DC voltage
Calibrator utilizes innovative technology to deliver
ultra-precision, ultra-stable DC Voltage from
100 nanovolts to 1200 Volts. Designed to meet the most
critical calibration laboratory requirements, the 2701C
is at home on the production line or in the field. This
lightweight, rugged, line powered instrument requires
only 15 seconds warm-up.
Performances
That Lead the Way
When something
really outstanding comes along, it's going to gain a lot
of attention. That's been the case of the 2701C since we
originally introduced it back in 1979. We're flattered
that other instrument manufactures have tried to imitate
our innovative features and design technologies. But,
while others have been busy trying to imitate the 2701C,
we've been busy improving it. Dozens of design
enhancements and three models later, we're proud to
introduce the 2701C with standard features like:
- "Cover-on"
Automatic Calibration
- 100%
Overrange
- Automatic
Bi-polar Output
- Built-in
Kilovolt Amplifier
- One Touch
Local and Remote Sensing
- 200mV Full
Scale Divided Output Range
- All Outputs
available through Two Terminals
- Basic
Accuracy Specified ± 5°C from Cal
Temp
- 25mA Current
Sourcing All Active Range
- Micro-processor Enhanced Reliability
- 5 Voltage
Ranges (200mV through 1200V)
- Front and
Rear Terminals
- Illuminated
High Voltage Warning
- 0.5 PPM
Resolution
Innovation in
Attenuation
The digital
attenuator is the design key to long term stability,
reliability and interfaceability. Because the crystal
controlled attenuator is digital, it can't drift like the
conventional Kelvin Varley divider approach. Also, the
switch contact resistance no longer becomes a maintenance
headache as it is removed from the accuracy determining
circuit.
Ultra-Low Noise
2701LC Version Available
For
applications such as analog to digital converter testing,
where low noise output is extremely desirable, option
"LNF" provides this enhancement to the Valhalla
2701C. Deleting the kilovolt amplifier provides a major
improvement in low noise performance. Additionally, by
substituting a toroidal transformer stage with 40VDC
maximum output, a significant noise reduction is attained
while maintaining a low level safe output.
0.1 Micro Volt
to 1.2KV Direct Output-Standard
The 2701C have
a built-in kilovolt amplifier that delivers up to
1200.000 volts with a full 25mA drive current capability.
The 2701C is also short-circuit damage proof on all
ranges, and features a 200mV divided output range with
100 nanovolts resolution.
System
Capability
Designed from
the ground up for systems interfaceability, the 2701C
offers unmatched user convenience for system
installation. To start with, all output stimuli are
available from two terminals on the front and rear
panels, so once you've plugged in there's no fumbling
around to get to separate divider or high voltage
connector. Also, if remote sensing is required a touch of
a button engages the 4-wire output mode. The optional
microprocessor controlled IEE-488 talk/listen interface,
"TL-3", is extremely user-friendly, and provides a
six-channel relay driver port for remote range selection
on the 2500EP AC-DC Current Calibrator.
SPECIFICATIONS
Accuracy
| PPM OF SETTING ±
RANGE NOISE |
|
From
Factory |
1
Year |
| 200mV |
±
20 ppm ± 1 V |
±
30 ppm ± 2 V |
| 2V
|
±
15 ppm ± 4 V |
±
25 ppm ± 6 V |
| 20V
|
±
13 ppm ± 13 V |
±
22 ppm ± 50 V |
| 120V |
±
14 ppm ± 250 V |
±
23 ppm ± 400 V |
| 1200V |
±
15 ppm ± 2.5 V |
±
24 ppm ± 4 V |
Noise &
Stability
| RANGE |
NOISE (0.1
TO 10HZ) |
24HR
STABILITY (DC TO 0.2HZ) |
| 200mV |
1 V |
1 V |
| 2V
|
2 V |
±
0.5 ppm ± 2 V |
| 20V
|
15 V |
±
0.5 ppm ± 10 V |
| 120V |
150 V |
±
0.5 ppm ± 100 V |
| 1200V |
1.5mV |
±
0.5 ppm ± 1mV |
Temperature
Coefficient
The temperature
coefficient adder is applicable only when outside of the
± 5°C window of the calibration
temperature.
| RANGE |
PPM OF
SETTING + V/1°C |
| 200mV |
2.5 ppm +
0.1 |
| 2V
|
1.5ppm +
0.6 |
| 20V
|
1.5ppm +
6 |
| 120V |
1.5ppm +
30 |
| 1200V |
1.5ppm +
300 |
General
Performance Characteristics
The table below
gives the general operating characteristics of the 2701C.
Resolution refers to the minimum adjustment that
may be made in the selected range. Max Current
refers to the maximum amount of current that may be
supplied by the selected range without affecting its
accuracy (4-wire mode must be used). Wideband
Noise refers to the maximum amount of microprocessor
and power supply noise (from 10Hz to 10kHz) that may be
present on the output voltage. The Linearity of
the 2701C is defined as the maximum allowable deviation
from a straight line between the zero and full-scale
outputs on each range. Output Impedance refers to
the impedance of the voltage amplifier.
| 2701C PERFORMANCE
CHARACTERISTICS |
| Range |
Resolution |
Max Current |
Wideband Noise |
Linearity |
Output Impedance |
| 200mV |
100nV |
--- |
25 VRMS
|
±
0.5 V |
450 |
| 2V
|
1 V |
25mA |
80 VRMS
|
±
1 V |
1m |
| 20V
|
10 V |
25mA |
130 VRMS
|
±
10 V |
5m |
| 120V |
100 V |
25mA |
500 VRMS
|
±
100 V |
50m |
| 1200V |
1mV
|
25mA |
800 VRMS
|
±
1mV |
500m |
Output Settling
Time
The table below
gives the maximum allowable error remaining after the
given time following a change in output voltage. The
errors are given in ppm of output voltage
change.
| RANGE |
0.5
SECONDS |
1
SECOND |
10
SECONDS |
| 20V and
below |
20
ppm |
5
ppm |
1
ppm |
| 120V |
30
ppm |
7
ppm |
2
ppm |
| 1200V |
50
ppm |
10
ppm |
5
ppm |
Other
Specifications
Warm-up
Time: 15 seconds to within 15 ppm of final value 30
minutes to rated specifications
Power: 115/230V ± 10% at 45 to 65Hz at 80VA
Size: 89mm (3.5") high x 432mm (17") wide x 432mm (17")
deep
Weight: 11Kg (24lbs) net, 13Kg (29lbs) shipping
Operating Temperature Range: 0 to 50°C
Storage: -30°C to 70°C
Humidity: 70% RH max at 40°C
(non-condensing)
Option LNF
Specifications
The Model 2701C is available
with a modification that reduces the Wideband noise. This
modification is designed as Option "LNF". A 2701C
equipped with Option LNF has a maximum output voltage of
± 40 volts. The following Wideband Noise
specifications apply to instruments equipped with Option
LNF only.
| RANGE |
10HZ TO
1KHZ |
1KHZ TO
10KHZ |
| 200mV |
10 VRMS
|
5 VRMS
|
| 2V
|
10 VRMS
|
10 VRMS
|
| 20V
|
15 VRMS
|
10 VRMS
|
| 120V |
100 VRMS
|
25 VRMS
|
|