Logo    Valhalla Model 4176 Specifications

Range Characteristics and Accuracy

Range Full Scale Resolution Test Current Accuracy Temperature Coefficient

20mΩ 20.000mΩ 1μΩ 1A ±0.02% of reading ±0.02% of range ±0.002%/°C

.2Ω .30000mΩ 10μΩ 1A ±0.02% of reading ±0.02% of range ±0.002%/°C

3.0000Ω 100μΩ 100mA ±0.02% of reading ±0.02% of range ±0.002%/°C

20Ω 30.000Ω 1mΩ 10mA ±0.02% of reading ±0.02% of range ±0.002%/°C

200Ω 300.00Ω 10mΩ 100μA ±0.02% of reading ±0.02% of range ±0.002%/°C

2kΩ 3.0000kΩ 100mΩ 1A ±0.02% of reading ±0.02% of range ±0.002%/°C

20kΩ 30.000kΩ 10Aμ ±0.02% of reading ±0.02% of range ±0.002%/°C


Temperature Compensation Mode Accuracy

Range For T<25°C For T>25°C

All Ranges ±.02% of rdg ±.02% of rng ±.001% of (25°C - T) ±.02% of rdg ±.02% of rng ±.001% of (T - 25°C )


General Specifications
Dispaly Type: VFD
A/D Conversion Rate: 45 conversions/second
Display Update: programmable (100ms, 200ms, 300ms, 400ms, 500ms)
Overload:
20mΩ: 99.95% of range
200mΩ thru 20kΩ: 119.95% of range
Overload Indication: flashes "OVERLOAD"
Terminal Configuration: 4-wire Kelvin
Test Current Polarity: Positive (flows Hi to Lo)
Test Current Compliance Voltage: 5V minimum
Settling Time: 300msec

General Specifications
Dispaly Type: VFD
A/D Conversion Rate: 45 conversions/second
Display Update: programmable (100ms, 200ms, 300ms, 400ms, 500ms)
 

General Specifications
Overload Indication: flashes "OVERLOAD"
Terminal Configuration: 4-wire Kelvin
Test Current Polarity: Positive (flows Hi to Lo)