- Dynamic
Range: 0 to 200% of range up to maximum rated
output
- Maximum
Output Current: 100A DC or RMS AC
- DC Accuracy:
(180 days 25°C ± 5°C) all ranges
<100A ±0.015% of range ±0.03% of
output 100 Amp Range: ± .03% of range
±0.03% of output
- AC Accuracy:
(180 days)( 25°C ± 5°C) ±0.15% of
range ± 0.1% of output to 100Hz ±0.2% of
range ± 0.2% of output to 400Hz ±0.3% of
output ± 0.3% of range to 1KHz
- Input
Impedance: 100 Kilohms
- Input/Output
Ratio: 2.00000 volt input provides full scale output
(1.0000 volt input produces 100.0 amps
output)
- Stability:
±0.0005% - 15 min. (±0.005% for 24
Hours)
- Maximum
Input: 3 VDC or 2 VAC (RMS)
- Response
Time: 1 millisecond to ±0.01% of final value
following input amplitude or frequency
change
- Maximum
Isolation Voltage: ± 200 VDC or Peak
AC
- Input Common
Mode Rejection Ratio: 60db @ DC linearly decreasing to
40 db @ maximum frequency
- Load
Regulation: Output current changes less than ±
.002%
- Temperature
Range: 0°C to 50°C
- Temperature
Coefficient: ±0.001% of output ±0.002% of
range/°C below 20°C and above 30°C (double
above for temperature coefficient for AC)
- Compliance
Voltage: 4 Volts
- Size:
266mm/10?" L x 432mm/ 17" W x 584mm/23" D
- Weight: 45
Kg/100 lbs. Net 66 Kg/140 lbs. shipping wt.
- Power:
115/230 VAC ± 10% 50-60Hz 2000 watts
|