'Manufacturers of Digital Power Analyzers, Digital Wattmeters, Harmonic Power Analyzers, Wideband Spectrum Power Analyzers, ATC Ohmmeters, Micro-Ohmmeters, Milliohm-Meters, Digital Ohmmeters, Igniter Testers, Current Calibrators, AC-DC Active Current Shunts, Hipot Testers, Body Comp Scales and more…'
 
Applications
Igniter 								Testers
Wattmeters
Power Analyzers
Ohmmeters
High Voltage
Calibration
Body Fat Scale
Options
Accessories
Calibration Procedures
Manuals
Specifications
History
Technical Support
Contact Us
Home
 
Option-10 - 0-3000A Flexible Current Clamp
Opt-10 image
Option-10 is an AC current probe utilising the Rogowski principle. It can be used to measure currents up to 3000A when used in conjunction with 2400 Series Power Analyzers. The flexible and lightweight measuring head allows quick and easy installation in hard to reach areas.


SPECIFICATIONS:

Current Ranges:

30 A / 300 A / 3000 A / AC RMS

Output Sensitivity (AC coupled):

100 mV / 10mV / 1mV per A

Load Impedance:

100 kOhm minimum

Accuracy (at +25°C):

± 1% of range (45 – 65 Hz)

Linearity (10% to 100% of range):

± 0.2% of reading

Noise:

8mV AC RMS (30 A),
2 mV AC RMS (300/3000A)

Frequency Range:

10 Hz to 50 kHz (- 3 dB)

Phase Error:

smaller ± 1º (45 – 65Hz),
± 10º (at 20 kHz)

Position Sensitivity (with cable > 25mm from the coupling):

± 2% of range

External field (with cable > 200mm from the head):

± 1% of range

Power Supply:

2 x AA MN 1500 LR6 alkaline
400 Hours, low battery indicator
or dedicated external power supply

Temperature Coefficient:

± 0.08% of reading per °C

Working Voltage (see Safety Standards section):

600 V ACRMS or DC


GENERAL CHARACTERISTICS:

Head Cable length…(double insulated):

610mm (24”) RR3030
915mm (36”) RR3030/36
1220mm (48”) RR3030/48

Head Cable Diameter:

14.3 mm

Head Bend Radius:

38.1 mm

Cable length (head to electronics):

2 meters

Output Connection:

Safety connector to clamp input of Power Analyzer (1 per phase)

Operating Temperature Range:

-20 to +90 °C (head),
-20 to +85 °C (electronics)

Storage Temperature Range:

-40 to +105 °C (head)
-20 to +85 °C (electronics)

Operating Humidity:

15% to 85% (non condensing)

Weight:

180 g (head), 190 g (electronics)

Safety Standards:

BSEN61010-1: 1993 and Amendment A2 : July 1995
Use of the probe on uninsulated conductors is limited to 600 V ACRMS or DC and frequencies below 1 kHz.

EMC Standards:

EN 61326 : 1998



Dimensions