Valhalla Models 2100 and 2101 Low Cost Benchtop
Power Analyzers
|

Valhalla
Scientific's 2101 Digital Power Analyzer |
Valhalla 2100
Series Power Analyzer Features Include:
- Up to 20
Amps/phase direct (self contained shunt
0.1%)
- Expandable
to 1000 amps (optional,see I-1000 C.T.)
- True Power Measurements, VIcos Ř
- Optional isolated GPIB I/O
- High Accuracy Measurement: 0.15% DC to 5KHz
- Bandwidth~ DC, 20 Hz to 50 KHz
- Zero to Unity Power Factor Response
- Accurate Regardless of Waveform Distortion
- Certificate
of N.I.S.T. traceabilty included at no extra
charge
|
|
 |
 |
|
The Valhalla
2100 and 2101 Digital Power Analyzers (V-A-W displays)
These two
benchtop single-phase Digital Power Analyzers are fully
loaded with measurement features you'd expect on
instruments costing three times the price. Both units
boast twin high resolution 4 1/2 digital displays, DC to
50 KHz frequency response, true power measurement, true
RMS voltage and current and built-in peak overload
indicators. The only difference between the two units are
their respective voltage ranges. The 2100 offers 600, 300
and 150 volt ranges while the 2101 (pictured above)
features 300, 150 and 30 volt ranges.
The Valhalla
2100 series power meters are also great measurement
instruments for college and university motor test
laboratories.
|
|
|
| Specifications |
 |
- Accuracy:
(Voltage and Current True RMS) DC & 20 Hz to 5 KHz:
± 0.1% of reading ± 6 counts
- 5 KHz to 15
KHz: ± 0.5 % of reading ± 6
counts
- 15KHz to 20
KHz: ± 0.75% of rdg ± 6 counts - useable
to 50 KHz with typically 1 % error per 10
KHz
- Maximum
Input: 600 VDC or RMS AC, 1500 V peak
- Displays:
Dual 4± Digit (20,000 counts) 7 segment
LEDs
- Watts
Accuracy: 25 °C ± 5°C, 1
year
- DC & 20 Hz
to 5 KHz: ± 0.25 % of reading ± 6
counts
- 5 KHz to 10
KHz: ± 0.5% of reading ± 0.5 % of
range
- 10 KHz to
20 KHz: ± 1% of reading ± 1% of
range
- 20 KHz to
50 KHz: additional 1% per 10 KHz above 20
KHz
- Power
Factor Response: Unity (1.000) to zero
(0.000)
- Power
Factor Accuracy: for inputs > than 10 % of full
scale
- DC to 1
KHz: ± 5 counts
- 1 KHz to 5
KHz: ± 10 counts
- 5 KHz to 10
KHz: ± 15 counts
- Extended
current capacity is available using 150 amp and 1000
amp clamp-on current transformers (option"I-150 and
option"I-1000"), low current 20 mA versions and low
power factor versions (low watts amplified x 10) are
also available.
- Displays:
Large 4 1/2 Digit (20,000 counts) 7 segment L.E.D.
type
|
|
|
Accessories
The following
accessories are compatible with 2100 Series Digital Power
Analyzers:
|
- Current
transformers with a 1000:1 step down ratio (1%
accuracy) include option "I-1000" (1000 amps capacity)
and option "I-150" (150 amp capacity)
- Rack Mount
Adapters -For 19" Racks
- Load
Extension Cord option "X-21" for easy 120 VAC load
connections
- Clamp-on
Current Transformer Option "I-150": Extends AC current
Measurement capability on the 2100 and 2101 to 150 amps
RMS. The 1000:1 output ratio is 2% accurate from 50Hz
to 400Hz. The unit accomodates up to ?" diameter
conductors.
- Clamp-on
Current Transformer Option "I-1000": Extends AC current
measurement capability to 1000 amps RMS. The 1000:1
ratio CT is 2% accurate from 50Hz to 400Hz. Data Output
Port Option "X21": Approximately three feet in length
for each half, this convenient adaptor cord plugs
directly into a standard 115V AC power outlet and mates
with the 2100 or 2101 via heavy duty banana jacks.
Female half provides for quick and easy load connect or
disconnect capability.
- For low
current applications Option "20mA" provides a lower
current range of 20mA, includes 200mA and 2.0A ranges;
however, it deletes the higher 20 ampere
range.
|
| Ordering Information
|
| Model 2100/2101
|
Benchtop Power Analyzer |
| Option "I-1000"
|
1000A Current Clamp |
| Option "I-150"
|
150A Current Clamp |
| Option "X-21"
|
Load Extension Cord Option
|
| Option "20mA"
|
Low Current Option
|
| Option "R-4" |
19" Rack
Adapter |
| Option
"M-2100" |
Additional
User/Maintenance Manual |
|
|
BACK
TO LIST
|